상세 보기
Endoscopic Removal of a Proximally Migrated Metal Stent during Balloon Sweeping after Stent Trimming
- Cho, Nam Jun;
- Lee, Tae Hoon;
- Park, Sang-Heum;
- Lee, Han Min;
- Hyun, Kyung Hee;
- ... Chung, Il-Kwun;
- 외 2명
WEB OF SCIENCE
7SCOPUS
13초록
Placement of a self-expanding metal stent (SEMS) is an effective method for palliation of a malignant biliary obstruction. However, metal stents can cause various complications, including stent migration. Distally migrated metal stents, particularly covered SEMS, can be removed successfully in most cases. Stent trimming using argon plasma coagulation may be helpful in difficult cases despite conventional methods. However, no serious complications related to the trimming or remnant stent removal method have been reported due to the limited number of cases. In particular, proximal migration of a remnant fragmented metal stent after stent trimming followed by balloon sweeping has not been reported. We report an unusual case of proximal migration of a remnant metal stent during balloon sweeping following stent trimming by argon plasma coagulation. The remnant metal stent was successfully removed with rotation technique using a basket and revised endoscopically.
키워드
- 제목
- Endoscopic Removal of a Proximally Migrated Metal Stent during Balloon Sweeping after Stent Trimming
- 저자
- Cho, Nam Jun; Lee, Tae Hoon; Park, Sang-Heum; Lee, Han Min; Hyun, Kyung Hee; Lee, Suck-Ho; Chung, Il-Kwun; Kim, Sun-Joo
- 발행일
- 2013-07
- 유형
- Article
- 권
- 46
- 호
- 4
- 페이지
- 418 ~ 422
- 언어
- ENG
- 출판사
- 대한소화기내시경학회
- 발행국가
- 대한민국
- 분량
- 5 페이지
- ISSN
- E 2234-2443
P 2234-2400