FLUCTUATION OF INFUSION PRESSURE DURING MICROINCISION VITRECTOMY USING THE CONSTELLATION VISION SYSTEM

Citations

WEB OF SCIENCE

14
Citations

SCOPUS

17

초록

Purpose: To measure fluctuations in infusion pressure and intraocular pressure (IOP) during vitrectomy performed using a flow-based IOP control system. Methods: Using 3 vitrectomized porcine eyes, the authors simultaneously measured infusion pressure and IOP during vitreous cutting and aspiration and after extraction of operative instruments in 23-gauge and 25-gauge system. The measurements were performed with the "IOP control" setting turned on or off. The efficacy of valved cannula and a built-in "IOP control limit" module in attenuation of infusion pressure fluctuation was evaluated. Results: At set pressure of 30 mmHg and 60 mmHg, the mean infusion pressure levels were 43.7 mmHg and 78.7 mmHg in the vitreous cutting mode, 67.4 mmHg and 101.2 mmHg in the aspiration mode, and 72.8 mmHg and 115.8 mmHg after extraction of the operative instrument, respectively, when the 23-gauge system was used. Use of valved cannulas effectively attenuated fluctuations in both infusion pressure and IOP. When the IOP control limit setting was "on," the compensatory infusion pressure increase was markedly limited and similar to the set pressure level when the IOP control limit was set at Level 2. Similar results were obtained when a 25-gauge system was used. Conclusion: Infusion pressure increased markedly during vitrectomy using a flow-based IOP control system.

키워드

Alcon Constellation Vision Systemflow-based intraocular pressure control systeminfusion pressureintraocular pressureset pressure of the deviceVISUAL-FIELD DEFECTSMACULAR HOLE SURGERYPARS-PLANA VITRECTOMYFLUID-AIR EXCHANGEINTRAOCULAR-PRESSURERECOGNIZED COMPLICATIONCATARACT-EXTRACTIONISCHEMIC INFARCTIONSURGICAL-PROCEDURESCLOSED VITRECTOMY
제목
FLUCTUATION OF INFUSION PRESSURE DURING MICROINCISION VITRECTOMY USING THE CONSTELLATION VISION SYSTEM
저자
Kim, Yong JoonPark, Sun HoChoi, Kyung Seek
DOI
10.1097/IAE.0000000000000625
발행일
2015-12
유형
Article
저널명
Retina
35
12
페이지
2529 ~ 2536