초록

High-resolution anorectal manometry (HR-ARM) and high-definition anorectal manometry (HD-ARM) catheters have closely spaced water-perfused or solid state circumferentially-oriented pressure sensors that provide much better spatiotemporal pressurization than non-high resolution catheters. This is a comprehensive review of HR-ARM and HD-ARM anorectal manometry catheter systems, the methods for conducting, analyzing, and interpreting HR-ARM and HD-ARM, and a comparison of HR-ARM with non-high resolution anorectal manometry. Compared to non-high resolution techniques, HR-ARM and HD-ARM studies take less time and are easier to interpret. However, HR-ARM and HD-ARM catheters are more expensive and fragile and have a shorter lifespan. Further studies are needed to refine our understanding of normal values and to rigorously evaluate the incremental clinical utility of HR-ARM or HD-ARM compared to non-high resolution manometry.

키워드

Anal canal; Constipation; Fecal incontinence; Manometry; Physiology; INTERNAL ANAL-SPHINCTER; NORMAL VALUES; OBSTRUCTIVE DEFECATION; ESOPHAGEAL MANOMETRY; PRESSURE TOPOGRAPHY; FECAL INCONTINENCE; PERINEAL DESCENT; DISORDERS; DIAGNOSIS; CONTINENCE
제목
How to Perform and Interpret a High-resolution Anorectal Manometry Test
저자
Lee, Tae Hee; Bharucha, Adil E.
DOI
10.5056/jnm15168
발행일
2016-01-30
유형
Review
저널명
Journal of Neurogastroenterology and Motility (JNM)
권
22
호
1
페이지
46 ~ 59