금 나노입자를 이용한 새로운 당화혈색소의 검출 기술

Novel Detection Technology for Glycated Hemoglobin using Gold Nanoparticles

초록

We report a novel detection technology for glycated hemoglobin (HbA1c) that is measured primarily to identify the three-month average plasma glucose concentration. In enzymatic measuring of glycated hemoglobin, the generated hydrogen peroxide was then used as a reducing agent of gold (III) for the synthesis of gold (0). Gold nanoparticles obtained from this novel approach were measured by optical and piezoelectric methods. In optical method, we have developed polymer based film-type sensor cartridge filled with all the reagents for glycated hemoglobin analysis and the cartridge worked very well having the detection limit of 0.53% of glycated hemoglobin. On the other hand, quartz crystal microbalance (QCM) sensors also have been developed to determine the abilities of surface modified QCM sensors at various levels of the concentration of glycated hemoglobin to bind gold nanoparticles and limit of detection was 0.90%. Finally, despite of relatively lower sensitivities of QCM sensor and film-type optical sensor than well-plate based optical detection, these two sensors were available to measure the glycated hemoglobin level for diabetes patients and normal person

키워드

Glycated hemoglobin(HbA1c)Hydrogen peroxideGold nanoparticleFructosyl amino acid oxidase(FAOD)Film-type cartridgeQuartz Crystal Microbalance(QCM)
제목
금 나노입자를 이용한 새로운 당화혈색소의 검출 기술
제목 (타언어)
Novel Detection Technology for Glycated Hemoglobin using Gold Nanoparticles
저자
이수석
DOI
1225-5475/eISSN2093-7563
발행일
2016
저널명
센서학회지
25
6
페이지
435 ~ 439