상세 보기
A Comparison of Central Corneal Thickness Measurements and Measurement Repeatability Using Three Imaging Modalities
- Woo, Sang Earn;
- Lee, Si Hyung
WEB OF SCIENCE
0SCOPUS
0초록
Purpose: To compare and analyze central corneal thickness (CCT) measurements and measurement repeatability according to the corneal thickness obtained with anterior segment optical coherence tomography (AS-OCT), ultrasound pachymetry (USP), and dual Scheimpflug imaging (Galilei G4 (R)). Methods: CCT measurements obtained using AS-OCT, USP, and a dual rotating Scheimpflug camera in 120 eyes of 60 healthy subjects were compared and the repeatability was analyzed. Results: The average CCT measurements using AS-OCT, USP, and a dual rotating Scheimpflug camera were 511.23 +/- 37.16, 525.78 +/- 39.22, and 531.43 +/- 41.57 mu m, respectively. The CCT measurement using dual rotating Scheimpflug imaging was statistically significantly thicker than that measured using AS-OCT (p < 0.05). A significant linear correlation was observed among OCT, USP, and dual rotating Scheimpflug measurements (r > 0.900, p < 0.001). All three methods showed high repeatability (intraclass correlation coefficient >= 0.996) for intra-examiner CCT measurements. Conclusions: The CCT measurement results using the three methods were significantly correlated and showed high repeatability. However, the AS-OCT measurement was significantly thinner than that obtained using a dual rotating Scheimpflug camera. CCT measurement with a dual rotating Scheimpflug camera is a better alternative for USP than AS-OCT. Therefore, CCT should be interpreted in the context of the instrument used.
키워드
- 제목
- A Comparison of Central Corneal Thickness Measurements and Measurement Repeatability Using Three Imaging Modalities
- 제목 (타언어)
- 세 가지 기기를 이용하여 측정한 중심각막두께의 비교 및 반복성 비교
- 저자
- Woo, Sang Earn; Lee, Si Hyung
- 발행일
- 2021-02
- 유형
- Article
- 저널명
- 대한안과학회지
- 권
- 62
- 호
- 2
- 페이지
- 184 ~ 192