Evaluation of image and dose according to I-dose technique when performing a CT scan

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

In this study, we applied the iterative reconstruction technique to improve image quality (I-dose) and evaluated its usability by analyzing the quality of the resulting image and evaluating the dose. To perform the scans, we fixed the uniform module (CTP 486's section) 4 on the table of the computed tomography (CT) device with the American association of physicists in medicine (AAPM) phantom and located it in the center where the X-rays could be generated by using a razor beam. Then, we set up the conditions of 120 kilovoltage peak (kVp), 150 milliampere second (mAs), collimation 4 x 0.625 mm, and a standard YA (Y-Sharp) filter. Next, we formed two groups: Group A in which I-dose was not applied and Group B in which I-dose was applied. According to the rod in the middle, after fixing the location of (A) at 12 o'clock, (B) at 3 o'clock, (C) at 6 o'clock, and (D) at 9 o'clock to evaluate the image quality, the CT number was measured and the noise level was analyzed. Using the AAPM phantom with doses of 50, 100, 200, 250, and 300 mAs by 80, 100, and 120 kVp, a dose analysis was performed. After scanning, the CT numbers and noise level were measured 20 times as a function of the I-dose levels (1-7). After applying I-dose at 6, 9, 12, and 3 o'clock, when a higher I-dose was applied, a lower noise level was measured. As a result, it was found that when applying I-dose to the AAPM phantom, the higher the level of I-dose, the lower the level of noise. When applying I-dose, the dose can be reduced by 60%. When I-dose is applied when taking CT scans in a clinical study, it is possible to lower the dose and lower the noise level.

키워드

I-dosenoisedoseITERATIVE RECONSTRUCTION TECHNIQUECHEST CTCOMPUTED-TOMOGRAPHYREDUCTION STRATEGYQUALITYANGIOGRAPHYPROTOCOLSEXPOSUREPATIENT3D
제목
Evaluation of image and dose according to I-dose technique when performing a CT scan
저자
Ryu, S. W.Lee, H. K.Cho, J. H.
DOI
10.1080/10420150.2015.1038714
발행일
2015-06-03
유형
Article
저널명
Radiation Effects and Defects in Solids
170
6
페이지
490 ~ 500