Measurement reliability of automated oscillometric blood pressure monitor in the elderly with atrial fibrillation: a systematic review and meta-analysis

Citations

WEB OF SCIENCE

12
Citations

SCOPUS

14

초록

Objectives This study aimed to identify whether automated oscillometric blood pressure monitor (AOBPM) is a reliable blood pressure (BP) measurement tool in geriatric patients with atrial fibrillation (AF) with high variability in BP and to evaluate whether it can be applied in practice. Methods Electronic searches were performed in databases including MEDLINE, EMBASE, the Cochrane Library, and CINAHL by using the following keywords: 'atrial fibrillation,' 'atrial flutter, 'blood pressure monitor', 'sphygmomanometer.' The QUADAS-2 was applied to assess the internal validity of selected studies. Meta-analysis was performed using RevMan 5.3 program. Design: Systematic review. Results We identified 10 studies, including 938 geriatric patients with AF. We compared with the previously used BP measurement method (mainly office) and AOBPM, and the patients with AF were divided into the AF-AF (atrial fibrillation rhythm continued) and AF-SR groups (sinus rhythm recovered). The difference in the systolic BP was -3.0 mmHg [95% confidence interval (CI): -6.58 to 0.59] and -1.62 (95% CI: -6.08 to 2.84) mmHg in the AF-AF and AF-SR groups, respectively. The difference in the diastolic BP was 0.17 (95% CI: -2.90 to 3.25) mmHg and -0.23 (95% CI: -5.11 to 4.65) mmHg, respectively. Conclusion This review showed that the BP difference from AOBPM compared with the auscultatory BP method was less than 5 mmHg in the elderly with AF. This difference is acceptable in clinical practice. However, AOBPM compared with invasive arterial BP in the diastolic BP was a difference of 5 mmHg or more, and so its accuracy cannot be assured.

키워드

aged; atrial fibrillation; blood pressure determination; sphygmomanometer; systematic review; MEASURING DEVICES; ANEROID SPHYGMOMANOMETERS; HEART-ASSOCIATION; HYPERTENSION; MANAGEMENT; SOCIETY; HOME; RECOMMENDATIONS; VALIDATION; DIAGNOSIS
제목
Measurement reliability of automated oscillometric blood pressure monitor in the elderly with atrial fibrillation: a systematic review and meta-analysis
저자
Park, Seong-Hi; Choi, Yun-Kyoung
DOI
10.1097/MBP.0000000000000414
발행일
2020-02
유형
Review
저널명
Blood Pressure Monitoring
권
25
호
1
페이지
2 ~ 12