Setup deviations for whole-breast radiotherapy with TomoDirect: A comparison of weekly and biweekly image-guided protocols

  • Jung, Jae Hong
  • Jung, Joo-Young
  • Bae, Sun Hyun
  • Moon, Seong Kwon
  • Cho, Kwang Hwan
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

The purpose of this study was to compare patient setup deviations for different image-guided protocols (weekly vs. biweekly) that are used in TomoDirect three-dimensional conformal radiotherapy (TD-3DCRT) for whole-breast radiation therapy (WBRT). A total of 138 defined megavoltage computed tomography (MVCT) image sets from 46 breast cancer cases were divided into two groups based on the imaging acquisition times: weekly or biweekly. The mean error, three-dimensional setup displacement error (3D-error), systematic error (I ) pound, and random error (sigma) were calculated for each group. The 3D-errors were 4.29 +/- 1.11 mm and 5.02 +/- 1.85 mm for the weekly and biweekly groups, respectively; the biweekly error was 14.6% higher than the weekly error. The systematic errors in the roll angle and the x, y, and z directions were 0.48A degrees, 1.72 mm, 2.18 mm, and 1.85 mm for the weekly protocol and 0.21A degrees, 1.24 mm, 1.39 mm, and 1.85 mm for the biweekly protocol. Random errors in the roll angle and the x, y, and z directions were 25.7%, 40.6%, 40.0%, and 40.8% higher in the biweekly group than in the weekly group. For the x, y, and z directions, the distributions of the treatment frequency at less than 5 mm were 98.6%, 91.3%, and 94.2% in the weekly group and 94.2%, 89.9%, and 82.6% in the biweekly group. Moreover, the roll angles with 0 - 1A degrees were 79.7% and 89.9% in the weekly and the biweekly groups, respectively. Overall, the evaluation of setup deviations for the two protocols revealed no significant differences (p > 0.05). Reducing the frequency of MVCT imaging could have promising effects on imaging doses and machine times during treatment. However, the biweekly protocol was associated with increased random setup deviations in the treatment. We have demonstrated a biweekly protocol of TD-3DCRT for WBRT, and we anticipate that our method may provide an alternative approach for considering the uncertainties in the patient setup.

키워드

TomotherapyBreast cancerSetup errorMVCTImage protocolRADIATION-THERAPYTOMOTHERAPYIRRADIATIONCANCERMOTIONERROR
제목
Setup deviations for whole-breast radiotherapy with TomoDirect: A comparison of weekly and biweekly image-guided protocols
저자
Jung, Jae HongJung, Joo-YoungBae, Sun HyunMoon, Seong KwonCho, Kwang Hwan
DOI
10.3938/jkps.69.1247
발행일
2016-10
유형
Article
저널명
Journal of the Korean Physical Society
69
7
페이지
1247 ~ 1253