Improving Accuracy and Reliability of Hearing Tests: Measurement Standards for Audiometric Devices

  • Cho, Wan-Ho
  • Lee, Jihyun
  • Seo, Young Joon
  • Kong, Tae Hoon
  • Suh, Michelle J.
  • ... Choi, Seong Jun
  • 외 5명
Citations

WEB OF SCIENCE

5
Citations

SCOPUS

5

초록

Pure-tone audiometry, using an audiometer, is the fundamental hearing test for diagnosing hearing loss. The requirements of the devices and the detailed process for calibrating the related equipment are described in international standards. However, traceable calibration and uncertainty evaluation processes are not widely accepted or applied to the qualification and maintenance of audiometric equipment. Here, we briefly review standard measurement systems for audiometric devices and introduce their calibration procedures. The uncertainty of each calibration process was investigated, and its impact on hearing test results was considered. Our findings show that the traceability of each procedure can be secured, satisfying the uncertainty requirement and being sufficiently smaller than the permissible deviation from the audiometer requirement. To guarantee the objectivity and reliability of hearing tests and maintain low uncertainty, close cooperation and mutual understanding between the metrology field and the medical community are necessary. J Audiol Otol 2024;28(3):167-175

키워드

AudiometryMeasurement standardsTraceabilityUncertaintyCalibration
제목
Improving Accuracy and Reliability of Hearing Tests: Measurement Standards for Audiometric Devices
저자
Cho, Wan-HoLee, JihyunSeo, Young JoonKong, Tae HoonSuh, Michelle J.Jin, In-KiOh, Soo HeeLee, Hyo-JeongChoi, Seong JunPark, Kyung-HoCha, Dongchul
DOI
10.7874/jao.2024.00227
발행일
2024-07
유형
Review
저널명
Journal of Audiology & Otology
28
3
페이지
167 ~ 175