Comparative Results of Vocal Fold Injury According to the Fiber-Based Laser in a Rabbit Vocal Fold Model

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

1

초록

Objectives. We assessed the safety of flexible fiber-based lasers including the potassium-titanylphosphate (KTP), CO2, and Holmium lasers using a rabbit vocal fold model and provided the safety evidence of laser before human clinical trial. Methods. In all, 120 male New Zealand white rabbits were used. Each laser was used to induce acute and chronic vocal fold injuries in 40 rabbits. The same laser energy with the same intensity and frequency were used in all cases, and we evaluated outcomes via surface scanning electron microscopy (SEM) and histological examination 1 day after injury. Histological and high-speed vocal fold vibration examinations were evaluated 1 month after injury. Surface injury roughness grading was done via SEM, and the Acute injury ratio and lamina propria (LP) ratio were also calculated. The dynamic glottal gap was measured through functional analyses based on recordings from a high-speed digital camera. Results. The Holmium laser caused significantly more vocal fold damage than did the KTP and CO2 lasers as revealed by SEM and evaluation of acute and chronic injuries. Functional analysis with high-speed digital camera indicated that the holmium laser reduced dynamic glottal gap compared with the normal vocal fold, where the other lasers did not. Conclusions. The histological and functional analyses of rabbit vocal fold experiments indicated that fiberbased laryngeal laser surgery for vocal fold lesions could be performed relatively safely using a KTP or CO2 laser.

키워드

Animal modelFiber-basedLaserVocal foldFIBROBLAST-GROWTH-FACTORINJECTIONSCAR
제목
Comparative Results of Vocal Fold Injury According to the Fiber-Based Laser in a Rabbit Vocal Fold Model
저자
Jeon, Gyeong HwaSeon, Sang WooLee, Seung Won
DOI
10.1016/j.jvoice.2023.01.029
발행일
2025-07
유형
Article
저널명
Journal of Voice
39
4
페이지
1014 ~ 1018