Effects of NEX on SNR and Artifacts in Parallel MR Images Acquired using Reference Scan

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

The aim of this study was to investigate effects of the number of acquisitions (NEX) on signal-to-noise (SNR) and artifacts in SENSE parallel imaging of magnetic resonance imaging (MRI). 3.0T MR System, 8 Channel sensitivity encoding (SENSE) head coils were used along with an in-vivo phantom. Reference sequence of 3D fast field echo (FFE) was consisted of NEX values of 2, 4, 6, 8, 10 and 12. The T2 turbo spin echo (TSE) sequence used for exams achieved SENSE factors of 1.2, 1.5, 1.8, 2.0, 2.2, 2.5, 2.8, 3.0, 3.2, 3.5, 3.8 and 4.0. Exams were conducted five times for each SENSE factor to measure signal intensity of the object, the posterior phase-encode direction and frequency direction. And SNR was calculated using mean values. SENSE artifacts were identified as background signal intensity in the phase-encoded direction using MRIcro. It was found that SNR increased but SENSE artifacts reduced with NEX of 4, 8 and 12 when the NEX increased in reference scan. It is therefore concluded that image quality can be improved with NEX of 4, 8 and 12 for reference scanning.

키워드

reference scan; NEX; SNR; SENSE artifact; SENSE
제목
Effects of NEX on SNR and Artifacts in Parallel MR Images Acquired using Reference Scan
저자
Heo, Yeong-Cheol; Lee, Hae-Kag; Cho, Jae-Hwan
DOI
10.4283/JMAG.2013.18.4.422
발행일
2013-12
유형
Article
저널명
Journal of Magnetics
권
18
호
4
페이지
422 ~ 427